Troubleshooting

ArtefactCauseSolution
Muscle tremorsPatient tenseAsk the patient to relax and reassure
 Patient is coldEnsure that room is warm
Provide warming devices eg) blankets
 ParkinsonsMove limb electrodes closer to torso
Advise the patient to place their hands under their body
Wandering baselinePatient too hot (sweaty)Cool room
Thorough skin preparation with alcohol wipe and reapplication of electrodes
 Cable movementEnsure that the cables are adequately supported
 Respiratory swingAsk the patient to hold their breath
Readjust patient to more upright position
AC Mains interferenceFracture of wire within cableCheck leads and change if necessary
 Other electrical equipment nearbySwitch off all non-essential equipment